M106

An 'Abs/%T/Conc' spectrophotometer for under £900

M106 Student/QC Spectrophotometer

M106 Brochure - Click to Download
M106 Brochure

  • Philips Tungsten-Halogen source lamp
  • Hamamatsu Photodiode Detector
  • 1200 lines/mm grating monochromator
  • Wavelength range 335 to 1000nm
  • Narrow (8nm) bandpass
  • Automatic stray light filter selection
  • Auto-zero by push-button
  • Stable, noise-free readings
  • Read-out in Abs, %T, Concentration and Factor
  • RS232 output for printer or PC
  • PC software option
  • Holder for test-tubes, standard 10mm cuvettes
  • Accepts 1ml semi-micro cuvettes
  • COD vial holder as accessory

The fundamental reliability of the instrument is assured by the absence of motor drives.

This robustly built spectrophotometer combines a surprisingly high level of performance with a budget price. Wavelength is conveniently set using the rotary control and the instrument is ready for use within just a few minutes from switch-on. A membrane keyboard positioned next to the large, clear LCD display allows the operator to rapidly switch between %T, Abs and Concentration modes, and to set Zero Abs. The narrow 8nm bandpass is ideally matched to the wavelength range of 335 to 1000nm.

The remarkable stability and accuracy of the instrument will be welcomed by quality control and teaching laboratories. The sample compartment accepts either standard 10mm rectangular cuvettes, including disposable semi-micro cuvettes, or test-tubes, and the RS232 output provides a convenient link to a digital printer for hardcopy or a PC. A low-cost application program, which runs under Windows 9x includes the facility to create calibration curves on screen.

Technical Specifications

Wavelength Range
335 to 1000nm
Photometric accuracy
+/- 1%A, 0-2A
Optics
Single beam
Light Source
Tungsten-Halogen lamp
Monochromator
1200 lines/mm
Detector
Photodiode
Bandpass
8nm
Data output
RS232 for printer or PC
Wavelength accuracy
+/- 2nm
Noise
< 0.001A@ 0 A
Wavelength repeatability
+/- 1nm
Zero drift
<0.003A/hr
Stray light
< 0.3%T @ 340nm
Weight
6Kg
Photometric ranges
0-2A,0-125%T,0-1999C and 0-1999F
Dimensions (W,D,H)
408 x 308 x 185mm

[top]

Range by Model

Single Beam
M105
(Student Visible Range)
M106
(QC / Student Visible Range)
M107
(High Specification Visible Range)
M108
Programmable Visible Spectrophotometer
New M508
Basic Programmable UV/Visible Spectrophotometer
M501
(UV / Visible Spectrophotometer)

Double Beam
M550
(UV / Visible Spectrophotometer)

Fibre Optics
M400PC
(Fibre Optic Diode Array Spectrophotometer)

Special Systems
M350N
System for Measuring NATO colours of paints and fabrics.
M550 SPF
System for measuring sun protection factor of fabrics.

Camspec Ltd.
11 High Street, Sawston
Cambridge
UK
CB22 3BG
tel: +44 (0) 1223 836971
fax: +44 (0) 1223 836414

Camspec is an ISO 9001:2000 company. All instruments fully comply with CE requirements for safety and ESD/EMC emission and susceptibility.

 