M350N System for measuring the colours of NATO paints and fabrics
The M350N system comprises a special version of the M350 computer controlled double beam scanning spectrophotometer with an integrating sphere accessory installed. Instead of scanning from 190nm to 1100nm, the M350N scans from 350nm to 1200nm.
A white standard and a black standard enable the reflectance values of the sample
to be accurately determined over the wavelength range. The PC software displays
the reflectance spectrum together with upper and lower NATO specification limits.
Specification of M350N system with sphere installed:
- Light source: Tungsten-Halogen
- Wavelength range: 350 to 1200nm
- Angular divergence of beam: < ± 5º
- Beam area at sample: 10 x 10mm
- Bandpass: 4nm
- Wavelength accuracy: ± 0.5nm
- Transmittance accuracy: 20%T ± 0.3, 10%T ± 0.2, 2%T ± 0.1
- Wavelength calibration: Checked using Didymium filter at 807.1nm traceable to NPL
- Reflectance calibration: Using white and black standards traceable to NIST
|
Range by Model
Single Beam
M105
(Student Visible Range)
M106
(QC / Student Visible Range)
M107
(High Specification Visible Range)
M108
Programmable Visible Spectrophotometer
New M508
Basic Programmable UV/Visible Spectrophotometer
M501
(UV / Visible Spectrophotometer)
Double Beam
M550
(UV / Visible Spectrophotometer)
Fibre Optics
M400PC
(Fibre Optic Diode Array Spectrophotometer)
Special Systems
M350N
System for Measuring NATO colours of paints and fabrics.
M550 SPF
System for measuring sun protection factor of fabrics.
|